logo
Phone (972) 438-3774
Irving, Texas

FLUKE 1587 INSULATION MULTIMETER (FLUKE 1587)

( Item in stock! )

 

 

Specifications
Voltage DC
Maximum Voltage:  1000 V
Accuracy:  1577: ±(0.2%+2)
 1587: ±(0.09%+2)
Maximum Resolution:  0.001V
Voltage AC
Maximum Voltage:  1000 V
Accuracy:  1577: ±(2%+3)
 1587: ±(2%+3)
AC Bandwidth:  1587: 5kHz*
 *with low pass filter; 3db @ 800 Hz
Maximum Resolution:  0.1 mV
Current DC
Maximum Amps:  400mA
Amps Accuracy:  1577: ±(1.0%+2)
 1587: ±(0.2%+2)
Maximum Resolution:  0.01 mA
Current AC
Maximum Amps:  400mA
Amps Accuracy:  1577: ±(2.0%+2)*
 1587: ±(1.5%+2)*
 * 1kHz bandwidth
Maximum Resolution:  0.01 mA
Resistance
Maximum Resistance:  50 MΩ
Accuracy:  1577: ±(1.2%+2)
 1587: ±(0.9%+2)
Maximum Resolution:  0.1Ω
Capacitance(1587 only)
Maximum Capacitance:  9,999 µF
Accuracy:  ±(1.2%+2)
Maximum Resolution:  1 nF
Frequency (1587 only)
Maximum Frequency:  100 kHz
Accuracy:  ±(0.1%+1)
Maximum Resolution:  0.01 Hz
Temperature Measurement (1587 only)
 -40.0 °C to 537 °C
 -40.0 °F to 998 °F
 excluding probe
Diode Test (1587 only)
Range:  6 V
Resolution:  1 mV
Accuracy:  ±(2 % + 1)

 

Insulation Test
Minimum Test Current @ 1kΩ/V:
 1mA
Test Voltages:
1577:  500, 1000 V
1587:  50, 100, 250, 500, 1000 V
Maximum Resistance per Test Voltage:
50V:  
100V:  
250V:  
500V:  
1000V:  1577: 600 MΩ
 1587: 2 GΩ
Maximum Resolution per Test Voltage:
50V:  0.01 MΩ
100V:  0.01 MΩ
250V:  0.1 MΩ
500V:  0.1 MΩ
1000V:  0.1 MΩ
Accuracy per Test Voltage:
50V:  ±(3 % + 5)
100V:  ±(3 % + 5)
250V:  ±(1.5 % + 5)
500V:  1577: ±(2.0 % + 5)
 1587: ±(1.5 % + 5)
1000V:  1577: ±(2.0 % + 5)
 1587: ±(1.5 % + 5) to 600 MΩ, ±(10 % + 3) above 600 MΩ

 

Environmental Specifications
Operating temperature
 –20 °C to +55 °C
Storage temperature
 -40°C to +60°C
Humidity (Without Condensation)
 0% - 95% (10 °C – 30 °C)
 0% - 75% (30 °C – 40 °C)
 0% - 40% (40 °C – 55 °C)
Operating Altitude
 2000 m

 

Safety Specifications
Overvoltage category
 Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V
Agency Approvals
 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V

 

Mechanical & General Specifications
Size
 203 x 100 x 50 mm (with holster)
Weight
 624 g
Warranty
 3 years
Battery life Alkaline
Meter use:  1000 hours
Insulation test use:  Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.
IP Rating
 IP40
FLUKE 1587 INSULATION MULTIMETER
Our Price : $699.95  ea
 

Manufacturers

Perry

 

Calibration Specialty based in the Dallas, Texas area provides professional aircraft calibration services, airline calibration services, airline tool calibration, avionics tool calibration, avionics equipment calibration, helicopter calibration services, Fluke mulitmeter repair, Starrett calibration services, Mitutoyo calibration services and on-site calibration repair services.